Second Derivative Spectrophotometric Method for Simultaneous Determination of Tamsulosin and Finasteride in Pharmaceutical Formulations

    Mahesh Nasare, J. Satish, Ashok Kumar, V. V. L. N Prasad, M. Sanayaima Huidrom, Prakash V. Diwan
    Image of study
    TLDR A method was created in 2012 to accurately and precisely measure tamsulosin and finasteride in medicine using second derivative spectrophotometric method.
    In 2012, a simple and reliable second derivative spectrophotometric method was developed and validated for the simultaneous estimation of tamsulosin and finasteride in bulk and pharmaceutical formulations. The quantitative determination of the drugs was carried out using the second derivative values measured at 239 nm and 232 nm for tamsulosin and finasteride respectively. The solutions of standard and the sample were prepared in methanol. The calibration graphs constructed at their wavelengths of determination were linear in the concentration range of 1-8 µg ml-1 for tamsulosin and 12.5-100 µg ml-1 for finasteride. The low relative standard deviation values indicate good precision and high recovery values indicate accuracy of the proposed method. The developed method was found to be simple, accurate, precise, specific, sensitive, and reproducible, and could be directly and easily applied to pharmaceutical dosage forms.